ALP AUTOCLAVE (GTU-MAR) APPOINTMENT

 

ALP AUTOCLAVE USAGE INSTRUCTIONS

 

1. Be sure that power switch which localizes in front of machine is turned on.

2. Be sure that drain valce is closed position.

3. Open the lid and control the water level inside the machine. If water level is not enough, pour the  DISTILLED water up to base plate level.

4. Load the object into the machine, close and lock the lid.

5. Press the STOP key.

6. Set program condition according your objects (temperature and time ).

 ***If you don’t know how to use the autoclave, please GET PROFESSIONAL HELP.

7. Press the START key.

8. Unloaded objects after autoclave process is completed (below 80 ᵒC in chamber).

9. Turn off the power switch.

10. After the dirty material autoclave, clean the autoclave machine properly.

 

 

 

CLICK HERE FOR APPOINTMENT CALENDAR.

 